DEP SOPs
 

Site index

Home

 

 

Created by:

 

4.1              DECONTAMINATION

 

4.1.1            General

 

      4.1.1.1      All equipment shall be cleaned in a controlled environment and transported to the field pre-cleaned and ready to use.

 

      4.1.1.2      All equipment must be immediately rinsed with tap water after use, even if it is to be field cleaned for other sites.

 

      4.1.1.3      Equipment that is only used once (i.e. not cleaned in the field) must be tagged with the sample location, returned to the in-house cleaning facility and cleaned in a controlled environment.

 

4.1.2            Reagents

 

      4.1.2.1      Detergents specified in this document refer to Liquinox (or equivalent) or Alconox (or equivalent).  See section 4.1.4.1 NOTE 3 for restrictions on use.

 

      4.1.2.2      Solvents

 

a.            The solvent used in routine cleaning procedures shall be pesticide grade or nanograde isopropanol. Pesticide grade and nanograde are synonymous.

      1.      Other solvents (i.e. acetone or methanol) may be used if they are pesticide grade.  However, methanol is more toxic to the environment and acetone may be an analyte of interest for volatile organics.

      2.      Acetone SHALL NOT be used if volatile organics are of interest.

 

b.            Precleaning heavily contaminated equipment (see section 4.1.4.1 NOTE 2.a) may be done with reagent grade acetone and hexane.

 

      4.1.2.3      Analyte-free water sources shall be subject to the following criteria:

 

a.            Analyte-free water is water in which all analytes of interest and all interferences are below method detection limits.

b.            This type of water shall be obtained from a reputable source and documentation shall be maintained to demonstrate reliability and purity of analyte-free water sources (i.e. results from equipment blanks).  As a general rule, the following types of water should be used:

      1.      Milli-Q (or equivalent) - suitable for all analyses

      2.      Organic-free - may be suitable for only VOCs and extractable organics

      3.      Deionized water - suitable for only inorganic analyses (metals, nutrients, etc.)

      4.      Distilled water - suitable for inorganics or microbiology

c.            Analyte-free water shall always be used for blank preparation and for the final in-house decontamination rinse. 

d.            Analyte-free water shall be transported to the field in containers of suitable construction (see section 4.1.10).

 

      4.1.2.4      Decontaminated field equipment shall be protected from environmental contamination by securely wrapping and sealing with one of the following:

a.            Aluminum foil - grocery store type is acceptable;

b.            Untreated butcher paper; or

c.            Clean, disposable plastic bags may be used if only inorganics are of concern OR the equipment is first wrapped in foil or butcher paper.

 

      4.1.2.5      Acids

            All acids used for cleaning shall be reagent grade or better:

a.            10% hydrochloric acid is prepared by mixing one part concentrated hydrochloric acid with 3 parts deionized water.

b.            10% nitric acid is prepared by mixing one part concentrated nitric acid with 5 parts deionized water.

            Prepare acid solutions by slowly adding the concentrated acid to water.

 

4.1.3  Decontamination/Cleaning Protocols - General Considerations

 

      4.1.3.1      All sampling equipment (bailers, lanyards, split spoons, etc.) that come in contact with the sample must be cleaned/decontaminated before use.  The procedures that are applicable to the majority of sampling equipment are listed in Section 4.1.4.1.  Protocols for other specialized equipment are outlined in Sections 4.1.4.2 through 4.1.11.  Unless justified through documentation (i.e. company-written protocols and analytical records) and historic data (i.e. absence of analytes of interest in equipment blanks), the protocols outlined in Sections 4.1.4 through 4.1.11 shall be followed WITHOUT modifications.

 

      4.1.3.2      Sufficient clean equipment should be transported to the field so that an entire study can be conducted without the need for field cleaning.

 

      4.1.3.3      Unless otherwise justified, all field sampling equipment shall be precleaned in-house (office, lab, or base of field operations) prior to arrival on-site.

 

      4.1.3.4      All cleaning shall be documented for each piece of field equipment.

a.            In-field decontamination shall be documented in the field records.  These records shall specify the type of equipment that is cleaned and the specific protocols that are used (reference to internal SOPs for the cleaning protocols is acceptable).

b.            In-house cleaning records must identify the type of equipment (i.e. teflon bailers, PVC pump tubing, etc.), the date it was cleaned, the protocol or SOP that was used and the person who cleaned the equipment.

 

4.1.4            Decontamination/Cleaning Protocols - Sampling Equipment

 

      4.1.4.1  General Cleaning Procedure for Teflon and Stainless Steel Sampling Equipment

 

      This procedure shall be used when sampling for ALL parameter groups:  extractable organics, metals, nutrients, etc.) or if a single decontamination protocol is desired to clean all Teflon and stainless steel equipment.

      The cleaning procedures described below are for IN-FIELD cleaning, in-house cleaning must follow NOTE 1 of this section.

a.            Clean with tap water and lab grade soap (Liquinox or equivalent) using a brush, if necessary, to remove particulate matter or surface film (see NOTES 1, 2 and 3 of this section).

b.            Rinse thoroughly with tap water.

c.            If trace metals are to be sampled rinse with 10-15% reagent grade nitric acid (HNO3). The acid rinse should not be used on steel sampling equipment (bailers, augers, trowels, etc.).  See NOTE 4.

d.            Rinse thoroughly with deionized water (DI).  Enough water shall be used to ensure that all equipment surfaces are flushed with water.

e.            Rinse twice with isopropanol.  One rinse may be used AS LONG AS all equipment surfaces are thoroughly wetted with free-flowing solvent.

f.            Rinse thoroughly with analyte-free (see 4.1.2.3) water and allow to air dry as long as possible.

g.            Clean sampling equipment shall be wrapped (if appropriate) in aluminum foil, or in untreated butcher paper to prevent contamination during storage or transport to the field.

h.            If no further sampling is to be performed, equipment must be rinsed with tap water immediately after use.

 

NOTES:

            1.            In house Protocols require the following:

a.            Protocols must include the use of HOT tap water and cleaning in a contaminant-free environment. 

b.            Analyte-free water must be used as a final rinse.

      2.      The DER QA Section does not recommend cleaning heavily contaminated equipment in the field.  Such rigorous cleaning procedures should be performed at the base of operations.  Cleaning at the base of operations or in the field require the following:

a.            Prerinse equipment using the following solvents in the order described:  acetone-hexane-acetone.  The solvent rinse(s) must precede the soap and water wash described in the first step (a. above).

b.            In extreme cases, it may be necessary to steam clean the field equipment before proceeding with step a. above.

c.            If the field equipment cannot be cleaned utilizing these procedures, it should be discarded, unless further cleaning with stronger solvents and/or oxidizing solutions are effective.

      3.      Liquinox (or equivalent) is recommended by EPA, although Alconox (or equivalent) may be substituted if nutrients are not sampled.

      4.      If sampling for nutrients, a 10-15% reagent grade hydrochloric acid (HCl) rinse should be used (except stainless steel equipment).  If BOTH metals and nutrients are to be sampled, the HCl rinse must replace the HNO3 rinse, or the HNO3 rinse must be followed by the HCl rinse.

      5.      Hot detergent solutions and water rinses are not required for in-field decontamination.

 

      4.1.4.2      Teflon and glass equipment used to collect aqueous and solid samples for trace organics and metals [including oil & grease, TRPH, and total organic halogens (TOX)].

      In-house cleaning - follow 4.1.4.1, see NOTES 1 and 2

      In-field cleaning - follow 4.1.4.1, see NOTES 2 and 5

 

      4.1.4.3      Stainless Steel or Other Metallic Equipment used to collect Trace Organics and Metals (includes oil and grease, TRPH, TOX)

      In-house cleaning - follow 4.1.4.1, see NOTES 1 and 2, delete acid rinse

      In-field cleaning - follow 4.1.4.1, see NOTES 2 and 5, delete acid rinse

 

4.1.4.4      Plastic equipment Used to Collect ONLY Trace Metals

      In-house cleaning - follow 4.1.4.1 see NOTE 1, delete solvent rinse

      In-field cleaning - follow 4.1.4.1 see NOTE 5, delete solvent rinse

 

4.1.4.5      All Equipment Used to Collect ONLY Nutrients (NH4, NO3, etc.) and Inorganic Anions (Cl-, SO4, F-, etc.)

      In-house cleaning - follow 4.1.4.1 see NOTES 1, 3, and 4; delete solvent rinse

      In-field cleaning - equipment may be rinsed with analyte-free water immediately after use, then rinsed several times with sample water from the next sample.

 

4.1.4.6      All Equipment Used to Collect ONLY Demands and other inorganic non-metals

      In-house cleaning - follow 4.1.4.1 see NOTE 1, delete solvent and acid rinses

      In-field cleaning - equipment may be rinsed with analyte-free water immediately after use, then rinsed several times with sample water from the next sample.

 

4.1.5            Automatic Wastewater Samplers, Sampling Trains and Bottles

 

      4.1.5.1      General - All ISCO and other automatic samplers shall be cleaned as follows:

a.            The exterior and accessible interior (excluding the waterproof timing mechanisms) portions of automatic samplers shall be washed with laboratory detergent (see NOTE 3 above) and rinsed with tap water.

b.            The face of the timing case mechanisms shall be cleaned with a clean, damp cloth.

c.            All tubing (sample intake and pump tubing) should be checked and changed on a regular basis.  Tubing shall be changed if it has become discolored (i.e. mold and algae) or if it has lost its elasticity.

d.            New precleaned Silastic pump tubing (see 4.1.7.2 below) shall be installed.

 

      4.1.5.2      ISCO (or equivalent) Automatic Sampler Rotary Funnel and Distributor

a.            Clean with hot water, laboratory detergent (see NOTE 3 above) and a brush.

b.            Rinse thoroughly with deionized water.

c.            Replace in sampler.

 

      4.1.5.3      ISCO (or equivalent) Sampler Metal Tube

            Clean as outlined in 4.1.7.3 below.

 

      4.1.5.4      All Automatic Sampler Headers

a.            Disassemble header and using a bottle brush, wash with hot water and phosphate-free laboratory detergent.

b.            Rinse thoroughly with deionized water.

c.            Reassemble header, let dry thoroughly and wrap with aluminum foil.

 

      4.1.5.5      Reusable Glass Composite Sample Containers

a.            If containers are used to collect samples that contain oil, grease or other hard to remove materials, it may be necessary to rinse the container several times with reagent-grade acetone before the detergent wash.  If material cannot be removed with acetone, the container must be discarded.

b.            Wash containers thoroughly with hot tap water and laboratory detergent (see NOTE 3 above).

c.            Rinse containers thoroughly with hot tap water.

d.            If metals are to be samples, rinse containers with at least 10 percent nitric acid (see NOTE 4 above).

e.            Rinse containers thoroughly with tap water.

f.            Rinse containers thoroughly with deionized water.

g.            If organics are to be sampled, rinse twice with solvent (isopropanol recommended) and allow to air dry for at least 24 hours.

h.            Cap with aluminum foil, Teflon film or the decontaminated teflon-lined lid.

I.            After using, rinse with tap water in the field, seal with aluminum foil to keep the interior of the container wet, and return to the laboratory or base of operations.

j.            Containers should not be recycled or reused if:

1.      Containers used to collect in-process (i.e. untreated or partially treated) wastewater samples at industrial facilities;

2.      Any bottle having a visible film, scale or discoloration that remains after the above protocols have been used; or

3.      Containers were used to collect samples at pesticide, herbicide or other chemical manufacturing facilities that produce toxic or noxious compounds.  Such containers shall be properly disposed of (preferably at the facility) at the conclusion of the sampling activities.

            If the containers described above are reused, no less than 10% of the cleaned containers must be checked for the analytes of interest BEFORE use.  If found to be contaminated (i.e. constituents of interest are found at MDL levels or higher), the containers SHALL BE DISCARDED.

 

      4.1.5.6      Plastic Reusable Composite Sample Containers

a.            Use the protocols outlined in 4.1.5.5 above but omit the solvent rinse.

b.            Restrictions on reused or recycled containers listed in 4.1.5.5.j above shall be followed.

 

      4.1.5.7      ISCO (or equivalent) Glass Sequential Sample Bottles for Automatic Sampler Based for Sequential Mode

            Glass sequential sample bottles to be used in collecting inorganic samples shall be cleaned using the following protocol:

a.            Rinse with 10 percent nitric acid.

b.            Rinse thoroughly with tap water.

c.            Wash in dishwasher at wash cycle, using laboratory detergent cycle, followed by tap and deionized rinse cycles.  Note:  containers may be manually washed using protocol followed in 4.1.5.5 above.  Solvent rinse should be deleted.

d.            Replace bottles in covered, automatic sampler base; cover with aluminum foil for storage.

e.            Rinse bottles in the field with tap water as soon as possible after sampling event.

 

      4.1.5.8      ISCO (or equivalent) Glass Sequential Samples Bottles (Automatic Sampler Base for Sequential Mode) to be used for Collecting Samples for Organic Compounds

a.            Use cleaning protocols outlined in 4.1.5.5 above or use 4.1.5.7 above adding a solvent rinse after 7.c.  Allow containers to air dry for at least 24 hours before use.

b.            Replace in covered automatic sampler base; and cover with aluminum foil for storage.

 

      4.1.5.9      Bottle Siphons Used to Transfer Samples from Composite Containers

a.            Rinse tubing with solvent and dry overnight in a drying oven.

b.            Cap ends with aluminum foil and/or teflon film for storage.

c.            Seal in plastic for storage and transport.

d.            Flush siphon thoroughly with sample before use.

 

      4.1.5.10      Reusable Teflon Composite Mixer Rods

a.            Follow protocols outlined in Section 4.1.4.1 above.

b.            Wrap in aluminum foil for storage.

 

4.1.6            Filtration Equipment

 

      4.1.6.1        Dissolved Constituents using In-Line, Molded and Disposable Filter Units

a.            Peristaltic pump

      1.      The peristaltic pump is cleaned per section 4.1.8.2.b "Pumps used for sampling"

      2.      The silastic pump tubing is cleaned per section 4.1.7.2

      3.      If Teflon tubing is used, it must be cleaned per 4.1.7.1

      4.      Other tubing types [high density polyethylene (HDPE), etc.] must be cleaned according to the appropriate protocol listed in 4.1.7

b.            Other equipment types (e.g. pressurized teflon bailer)

      1.      Other types of equipment that utilize in-line, molded and disposable filters shall follow the appropriate cleaning regimen specified in Sections 4.1.4.1 through 4.1.4.6.

c.            Note:  filtration units for dissolved metals shall follow the configuration and requirements outlined in Figure 4.1 and Table 4.1.

 

      4.1.6.2      Dissolved Constituents using Non-disposable Filtration Units (i.e. syringes, "tripod assembly", etc.)

a.            Proceed with steps 1 through 5 of section 4.1.4.1, assembling and applying pressure to the apparatus after each rinse step (water and acid) to drive rinse material through the porous filter holder in the bottom of the apparatus.

b.            Remove and clean any transfer tubing per appropriate cleaning protocols (see section 4.1.7).

c.            Assemble the unit and cap both the pressure inlet and sample discharge lines (or whole unit if a syringe) with aluminum foil to prevent contamination during storage.

      1.      Note:  if the unit is to be used to filter only inorganic constituents (i.e. metals, nutrients, etc.), the unit may be sealed in a plastic bag to prevent contamination.

 

4.1.7     Sample Tubing Decontamination

 

      4.1.7.1      Teflon Tubing

 

a.            New Tubing - if new tubing is used once and discarded, preclean as follows:

            1.            rinse outside of tubing with pesticide-grade solvent

            2.            flush inside of tubing with pesticide-grade solvent

            3.            dry overnight in drying oven or equivalent (zero air, nitrogen, etc.)

b.            Reused Tubing - tubing shall be transported to the field in precut, precleaned sections.  The description below was written for in-lab cleaning only, FIELD CLEANING IS NOT RECOMMENDED.  In-house cleaning shall follow these steps:

      1.      Exterior of the tubing must be cleaned first by soaking the Teflon tubing in hot, soapy water in a stainless steel sink (or equivalent non-contaminating material).  Use a brush to remove any particulates, if necessary.

      2.      Take a small bottle brush and clean the inside of the tubing ends where the barbs are to be inserted.

      3.      Rinse tubing exterior and ends liberally with tap water.

      4.      Rinse tubing surfaces and ends with nitric acid, tap water, isopropanol, and finally analyte-free water.

      5.      Place tubing on fresh aluminum foil.  Connect all of the pre-cut lengths of Teflon hose with Teflon inserts or barbs.

      6.      Cleaning configuration:

a.            Cleaning reagents (soapy water, acid, isopropanol, etc.) shall be placed in an appropriately cleaned container (2-liter glass jar is recommended)

b.            Place one end of the teflon tubing into the cleaning solution.

c.            Attach the other end of the teflon tubing set to the INFLUENT end of the peristaltic pump.

d.            The effluent from the peristaltic pump may be recycled by connecting tubing from the effluent to the glass jar with the cleaning reagents.

e.            Recycling as described above may be done for all reagents listed in 7 below EXCEPT the final isopropanol rinse and the final analyte-free water rinse.  Disconnect the tubing between the effluent end of the pump and the jar of cleaning reagents. 

            1.            Isopropanol should be containerized in a waste container for proper disposal;

            2.            Analyte-free water may be discarded down the drain.

            7.            Using the above configuration:

a.            Pump copious amounts of hot, soapy water through the connected lengths.

b.            Follow with tap water, 10% nitric acid, tap water, then isopropanol, and finally analyte-free water.

c.            During the nitric acid and solvent rinses, turn the pump off and allow the reagents to remain in the tubing for 15 minutes, then continue with the next step.  Pumping a liter (each) of the nitric acid and solvent should be sufficient, depending on the inside diameter of the hose.

d.            Leave the Teflon inserts or barbs between the pre-cut lengths and cap or connect the remaining ends.

      8.      After the interior has been cleaned as described above, the exterior shall be rinsed with analyte-free water. 

      9.      The connected lengths should then be wrapped in aluminum foil or untreated butcher paper and stored in a clean, dry area until use.  Documentation for this cleaning shall be noted in the organization cleaning records.

 

      4.1.7.2     Silastic Rubber Pump Tubing used in Automatic Samplers and other peristaltic pumps - this tubing need not be replaced if the sample does not contact the tubing or if the pump is used for only purging (i.e., not being used to collect samples).  Tubing must be changed on a regular basis if used for sampling:

a.            Flush tubing with hot tap water and lab-grade detergent solution

b.            Rinse thoroughly with hot tap water

c.            Rinse thoroughly with DI water

d.            If used to collected metals samples, the tubing shall be flushed with 1+5 nitric acid, followed by thorough rinsing with DI water

e.            Install tubing in peristaltic pump or automatic WW sampler

f.            Cap both ends with aluminum foil or equivalent

NOTE:  Tubing must be changed at specified frequencies as part the preventative maintenance on the equipment.

 

      4.1.7.3      Stainless Steel Tubing

      The exterior and interior of stainless steel tubing shall be cleaned as follows:

a.            lab-grade soap solution with a long, narrow bottle brush

b.            rinse with (hot) tap water

c.            rinse with DI water

d.            rinse twice with isopropanol (if organics are sampled)

e.            completely air dry or thoroughly rinse with analyte-free water

 

      4.1.7.4      Glass Tubing

a.            Use new glass tubing

b.            Rinse with isopropanol (if organics are of concern)

c.            Air dry for at least 24 hours

d.            Wrap in aluminum foil or untreated butcher paper to prevent contamination during storage

e.            Discard tubing after use

 

      4.1.7.5      Miscellaneous Non-Inert Tubing Types (tygon, rubber, HDPE, PVC, etc.)

a.            New Tubing

      1.      As a general rule, new tubing may be used without preliminary cleaning.

      2.      New tubing shall be protected from potential environmental contamination by wrapping in aluminum foil, sealing in plastic bags or in the original sealed packaging.

      3.      If new tubing is exposed to potential contamination, the exterior and interior shall be thoroughly rinsed with hot tap water followed by a thorough rinse with deionized water.

      4.      If new tubing is to be used to collect samples, the tubing shall be thoroughly rinsed with sample water (i.e. pump sample water though the tubing) before collecting samples.

b.            Reused Tubing

      1.      Flush tubing with soapy solution of hot tap water and laboratory detergent.

      2.      Rinse exterior and interior thoroughly with hot tap water.

      3.      Rinse exterior and interior thoroughly with deionized water.

      4.      If used to collected metals samples, the tubing shall be flushed with 10% nitric acid, followed by thorough rinsing with DI water

      5.      Wrap tubing and cap ends in aluminum foil and seal in plastic to prevent contamination during storage and transport.

 

4.1.8     Pumps

 

      4.1.8.1      Submersible pumps

a.            Pumps used for purging and sampling trace metals and/or organics

      1.      Construction of pump body and internal mechanisms (bladders, impellers, etc.), including seals and connections must follow Table 4.1

      2.      Choice of tubing material must follow Table 4.1

      3.      Pump exterior must be cleaned per section 4.1.4.1. 

      Note:  the solvent rinse shall be deleted if the pump body is constructed of plastic (i.e. ABS, PVC, etc.)

      4.      Pump internal cavity and mechanism must be cleaned as follows:

a.            If for purging only, then the pump must be completely flushed with potable water prior to purging the next well.

b.            If for purging and sampling, then it must be completely disassembled (if so designed) and decontaminated between each well.

c.            If the pump cannot be (practically) disassembled, then the internal cavity/mechanism must be cleaned by pumping copious amounts of lab-grade soap solution, tap water, and DI water.

      5.      Teflon tubing will be cleaned per section 4.1.7.1.

      6.      Stainless steel tubing will be cleaned per section 4.1.7.3.

      7.      Cleaning of non-inert tubing shall follow the appropriate protocols in section 4.1.7 above (NOTE: very few options exist for non-inert tubing to be used for purging and/or sampling for trace organics).

 

b.            Pumps used for all other constituents (nutrient/ inorganic anions)

            1.            Pump construction - no restrictions

      2.      Pump tubing material - no restrictions

      3.      Scrub the exterior of the pump with appropriate metal-, phosphate- or ammonia-free detergent solution

      4.      Rinse the exterior with tap water and deionized water

      5.      Rinse interior of pump and tubing by pumping tap or deionized water through the system using clean bucket or drum.

 

      4.1.8.2      Above Ground Pumps Used for Purging and Sampling

 

a.            Pumps used for purging only

      1.      Exterior of the pump must be free of oil and grease

      2.      Tubing choice must follow restrictions as specified in Table 4.1.

      3.      Tubing coming in contact with formation water shall be cleaned according to the appropriate protocol for construction materials specified in section 4.1.7.

 

b.            Pumps used for sampling

      1.      Exterior of pump must be cleaned with a detergent wash followed by tap and DI water rinses

      2.      Tubing choice must follow restrictions as specified in Table 4.1.

      3.      Tubing coming in contact with formation water shall be cleaned according to the appropriate protocol for construction materials specified in section 4.1.7.

 

4.1.9     Field Instruments and Drilling Equipment

 

      4.1.9.1      Field Instruments (tapes, meters, etc.)

a.            wipe down equipment body, probes, and cables with lab-grade detergent solution,

b.            rinse thoroughly with tap water,

c.            rinse thoroughly with DI water, and

d.            wrap equipment in aluminum foil, untreated butcher paper or plastic bags to eliminate potential environmental contamination.

 

An optional isopropanol rinse may be performed if equipment comes in contact with contaminated water, etc.

 

      4.1.9.2      Soil Boring Equipment

 

      This pertains only to equipment that is NOT used to collect samples.  Split spoons, bucket augers and other sampling devices must be cleaned per requirements listed in 4.1.4.1 or 4.1.4.3 above.

a.            The engine and power head, auger stems, bits and other associated equipment should be cleaned with a power washer, steam jenny or hand washed with a brush using detergent (no degreasers) to remove oil, grease, and hydraulic fluid from the exterior of the unit.

b.            Rinse thoroughly with tap water.

 

      [[4.1.9.3      Well Casing Cleaning]]

      These protocols are included and are meant as RECOMMENDATIONS for cleaning well casing and riser pipes.  Recommendations from other DER programs, if different or more stringent, SHALL BE FOLLOWED.

[[a.            The Department recommends only using casing that is designed for subsurface environmental groundwater monitoring.  PVC pipe that is designed for well casing shall be transported to the field in original packing boxes.

 

b.            Other PVC casing (for plumbing, etc. uses) shall be prepared for cleaning by sanding stencils (if present) on those portions of riser pipe that may come in contact with formation water.  The ink used for stenciling may contribute to or contaminate the real samples.  Casing that has been contaminated with grease, hydraulic fluid, petroleum fuel, etc. may require additional cleaning or deemed unusable.

 

c.            All casing and riser pipe should be cleaned just prior to installation using the following protocol:

      1.      Steam clean all casing and riser pipe.  Steam cleaning criteria shall meet the following: water pressure - 2500 psi, water temperature - 200F.

      2.      Rinse thoroughly with tap (potable) water.  This tap water must be free of the analytes of interest, in effect analyte-free.  Some potable water sources may have very low levels of contamination (e.g. benzene, trihalomethane, trichloroethane).  A potable water source should only be used if it is known to be free of those contaminants that are being investigated.]]

 

4.1.10            Analyte-Free Water Containers

 

      Analyte-free containers are defined by EPA to be glass containers.  The DER QA Section will allow these containers to be made of glass, Teflon, polypropylene, or high density polyethylene (HDPE).  It is strongly recommended that inert glass or Teflon be used for containerizing organic-free sources of water.  Polypropylene is a good second choice.  HDPE, though not recommended, is acceptable.  Analyte-free water should not be left in these containers for extended periods, especially HDPE.  These containers should be filled up for a single sampling event and then emptied at the end of the sampling day.  EPA's cleaning procedure for glass (Teflon and polypropylene) is as follows:

 

      4.1.10.1      New Containers

a.            Wash per instructions in 4.1.4.1 [(delete solvent rinse if plastic (HDPE or polypropylene) containers are being cleaned].

b.            Cap with Teflon film, aluminum foil or the bottle cap.  Note:  the bottle cap shall be equipped with a teflon liner.  Aluminum foil or teflon film may be used as liner material.

 

      4.1.10.2      Reused Containers

a.            Immediately after being emptied, cap with aluminum foil, teflon film or the container cap.

b.            Wash container exterior with lab-grade detergent solution and rinsed with DI water

c.            Rinse interior twice with isopropanol (delete if containers are plastic, see 4.1.10.1.a above)

d.            Rinse interior thoroughly with analyte-free water,

e.            Invert and allow to drain and dry

f.            Fill container with analyte-free water and cap tightly with aluminum foil, Teflon film or the container cap.  Note:  the bottle cap shall be equipped with a teflon liner.  Aluminum foil or teflon film may be used as liner material.

g.            Water shall not be stored for more than 3 days prior to a sampling trip.

 

4.1.11   Ice Chests and Shipping Containers

 

      4.1.11.1      Wash exterior and interior of all ice chests with laboratory detergent (see NOTE 3).

 

      4.1.11.2      Rinse with tap water and air dry before storage.

 

      4.1.11.3      If the ice chest becomes severely contaminated with concentrated waste or other toxic or hazardous materials, it should be cleaned as thoroughly as possible, rendered unusable, and properly disposed.