DEP SOPs
 

Site index

Home

 

 

Created by:

 

8.0              PREVENTIVE MAINTENANCE

 

      Preventive maintenance is the key ingredient to possessing analytical equipment that will produce reliable data over the life of the instrument.

 

      Responsibility for preventive maintenance lies with the analyst and supervisory personnel in charge of the monitoring equipment. The analytical staff must be dedicated to the implementation of the preventive  maintenance program and always watchful for signs that there is a need for maintenance activities. The analyst and supervisory personnel must be supported by vendor specialists or in-house experts that handle activities beyond simple repairs or maintenance.

 

      The Preventive Maintenance Program must consist of:

 

      1.   A written PM schedule;

      2.      Documentation of all maintenance and repairs (records must be kept in an easily accessible manner);

      3.   Vendor operation and maintenance manuals available for all instrumentation; and

      4.   A written contingency plan specifying that backup equipment will be maintained for all instrumentation or stating that sampling events will be postponed and current sample load be invalidated until repairs are accomplished.  If samples are sent to another laboratory the subject laboratory must have an approved CompQAP for the parameters of concern and the DER Project Manager must be notified if the analytical work is being performed under a Quality Assurance Project Plan.

 

      Table 8.1 identifies preventive maintenance activities by instrument type required by DER with recommended frequencies.  Please note that it may be necessary to perform activities more frequently depending on heavy workloads, sample types analyzed and/or instrument performance.  If the instrument manufacturer recommends more frequent or additional maintenance activities these shall also be incorporated into the facility maintenance program.


Table 8.1

 

PREVENTIVE MAINTENANCE ACTIVITIES

 

INSTRUMENT/ACTIVITY

 

 

FREQUENCY

 

 

 

 

 

 

AA SPECTROPHOTOMETER (FLAME)

 

 

 

 

 

 

Clean nebulizer                                                                                   

SA

 

 

 

Clean spectrophotometer quartz windows                                          

W

 

 

 

Burner head cleaned; check tubing, pump and lamps                        

D(1)

 

 

 

O rings checked                                                                                   

M(1)

 

 

 

Fine Tune Wavelength; Check optics                                                 

A

 

 

 

Check electronics                                                                               

A(3,4)

 

 

 

 

 

 

AA SPECTROPHOTOMETER (FURNACE)

 

 

 

 

 

Check graphite tubes                                                                         

D(1)

 

 

 

Flush autosampler tubing                                                                   

D

 

 

 

Replace graphite electrodes and Shrouds                                          

SA

 

 

 

Clean furnace housing and injector tip                                               

D

 

 

 

Check electronics                                                                               

A(3,4)

 

 

 

 

 

 

AA SPECTROPHOTOMETER (COLD VAPOR & HYDRIDE)

 

 

 

 

 

 

Flush tubing (automated systems)                                                     

D(1)

 

 

 

Check absorption cell for vitrification                                                  

D(1)

 

 

 

Replace or clean quartz cell                                                                

3

 

 

 

Check electronics                                                                               

A(3,4)

 

 

 

 

 

 

ICP

 

 

 

 

 

 

Clean and realign torch                                                                       

M

 

 

 

Clean nebulizer and spray chamber; Check peristaltic                      

W(1)

 

 

 

pump tubing and vacuum pump oil

 

 

 

 

Check entire optical system (mirrors, windows, etc.)                         

A(3,4)

 

 

 

Check water lines, torch compartment and gases                             

D

 

 

 

Check electronics (voltages, waveforms, etc.)                                   

SA

 

 

 

Check wavelength calibration and adjust as needed                         

SA

 

 

 

Run interference (interelement) standard                                           

SA

 

 

 

 

 

 

GAS CHROMATOGRAPHS

 

 

 

 

 

GENERAL

 

 

 

 

 

Check septa, cylinder gas pressure,                                                 

D

 

   

 

oxygen/moisture traps

 

 

 

 

Bake out injector body                                                                      

2

 

 

 

Check electronics (voltages, waveforms, etc.)                                 

Q(3,4)

 

 

 

Check GC temperature calibrations (injector, oven, detector)

Q

 

 

COLUMNS

 

 

 

 

 

Change glass sleeve inserts, shorten ends of columns, change glass wool plugs, check for leaks or replace

3

 

 

ELECTRON CAPTURE DETECTOR

 

 

 

 

 

Wipe Tests                                                                                       

SA

 

 

 

Hydrogen cleaning                                                                           

3

 

 

 

Returned to factory for cleaning and refoil                                       

3,4

 


Table 8.1, cont.

PREVENTIVE MAINTENANCE ACTIVITIES

 

 

INSTRUMENT/ACTIVITY

 

 

FREQUENCY

GAS CHROMATOGRAPHS, Cont.

 

 

 

 

 

FLAME IONIZATION DETECTOR

 

 

 

 

 

Clean                                                                                               

Q

 

 

Replace Flame Tip                                                                           

A

 

PHOTO IONIZATION DETECTOR

 

 

 

 

 

Clean lamp                                                                                      

M(1)

 

HALL ELECTROLYTIC CONDUCTIVITY DETECTOR

 

 

 

 

 

Replace resin, change solvent and clean conductivity cell

3,4

 

 

Change Ni tube                                                                                

Q

 

NITROGEN PHOSPHORUS DETECTOR

 

 

 

 

 

Clean                                                                                                

Q

 

MASS SPECTROMETER

 

 

 

 

 

Replace vacuum pump oil and change desiccant                           

A

 

 

Check ion source and analyzer (dismantle and clean, replace parts as needed)

Q

 

 

Check mechanicals (vacuum pumps, relays, gas pressures and flows)

Q

 

 

Check mass calibration w/ FC-43 (perfluorotributyl- amine)

D

 

PURGE AND TRAP

 

 

 

 

 

Clean sparger                                                                                  

W

 

 

Change Trap                                                                                     

A

 

 

Bake Trap                                                                                        

2

 

 

Check purge flow                                                                              

M

 

 

Check for leaks                                                                                

M

 

 

 

 

 

HIGH PRESSURE LIQUID CHROMATOGRAPHY

 

 

 

 

 

 

Gas lines checked for leaks                                                               

D

 

 

Clean mobile phase flow system with nitric acid                                

SA

 

 

Clean detector flow cells with nitric acid                                             

SA (3)

 

 

Clean injection valve                                                                            

A

 

 

Check solvent filters                                                                            

W

 

 

Check pumps seals and check valve assemblies (clean and replace as pressures & flows of mobile phase indicate)

D

 

 

Lubricate oil felts, if present                                                                

M

 

 

Lubricate post column reagent pumps and check valve assembly oil felts

M

 

 

 

 

INFRARED SPECTROPHOTOMETER

 

 

 

 

 

 

Clean instrument housing                                                                    

M

 

 

Change desiccant and clean cells                                                      

Q

 

 

Clean windows                                                                                   

M

 

 

 

 


Table 8.1, cont.

PREVENTIVE MAINTENANCE ACTIVITIES

 

INSTRUMENT/ACTIVITY

 

 

FREQUENCY

 

AUTOANALYZERS

 

 

 

 

 

 

Check for leaks, flush out system and clean up spills after use

D

 

 

 

Clean sample probe and check all tubing for wear and discoloration

M(1)

 

 

 

Clean optics                                                                                       

Q

 

 

 

Oil sample motor, lubricate gears, clean flow cell                              

SA

 

 

 

Clean pump rollers, platens and colorimeter filters

M

 

 

 

 

 

 

ION CHROMATOGRAPH

 

 

 

 

 

 

Check for leaks                                                                                   

D

 

 

 

Check all lines for wear and discoloration                                           

W(1)

 

 

 

Check pump pistons                                                                             

A

 

 

 

 

 

 

TOC ANALYZER

 

 

 

 

 

 

Change injection needle, clean injection port,                                     

M

 

 

 

 and change catalyst

 

 

 

 

Inspect combustion tube                                                                       

SA

 

 

 

 

 

 

RADIOCHEMISTRY

 

 

 

 

 

LOW BACKGROUND GAS PROPORTIONAL COUNTERS

 

 

 

 

 

Clean sample drawers and windows                                                  

W

 

 

LUCAS CELLS

 

 

 

 

 

Clean Windows                                                                                    

M

 

 

 

 

 

 

REFRIGERATORS, INCUBATORS, OVENS

 

 

 

 

 

 

Clean interior                                                                                         

M

 

 

 

Check thermometer temperature against certified thermometer or equivalent

A

 

 

 

 

 

 

ANALYTICAL BALANCES

 

 

 

 

 

Clean pan and compartment                                                                

D

 

 

 

Check with class S weights                                                                    

M

 

 

 

Manufacturer cleaning and calibration                                                  

A

 

 

 

 

 

 

AUTOCLAVES

 

 

 

 

 

 

Gaskets checked                                                                                  

W(1)

 

 

 

Timing mechanism checked                                                                 

SA

 

 

 

Clean interior                                                                                        

M

 

 

 

Sterilization indicator tape                                                                     

D

 

 

 

 

 

 

MICROSCOPES

 

 

 

 

 

 

Clean optics                                                                                         

M

 

 

 

 

 

 


Table 8.1, cont.

PREVENTIVE MAINTENANCE ACTIVITIES

 

INSTRUMENT/ACTIVITY

 

 

FREQUENCY

UV/VIS SPECTROPHOTOMETER

 

 

 

 

 

 

Lamp alignment checked                                                                     

3,4

 

 

Windows cleaned                                                                                 

M

 

 

Check and adjust photomultiplier sensitivity and wavelength resolution

A

 

 

Replace lamp                                                                                        

4

 

 

Clean sample compartment before and after each use                        

D

 

 

Check electronics                                                                                 

A(3,4)

 

 

Adjust baseline for smoothness through entire wavelength range

SA

 

 

Clean cuvettes after each use                                                              

D

 

 

 

 

TOX ANALYZER (TOTAL ORGANIC HALOGENS)

 

 

 

 

 

 

Clean titration cell; clean inlet and exit tube                                         

W

 

 

Clean pyrolysis tube, recoat electrodes                                                

5

 

 

 

 

 

pH AND ION SELECTIVE ELECTRODES

 

 

 

 

 

PROBE

 

 

 

 

 Check probe for cracks and proper levels of filling solution; check reference junction; clean electrode

D(1)

 

 

Check response time                                                                          

D

 

METER

 

 

 

 

Check batteries and electronics for loose connections and cracked leads

D(1)

 

 

 

 

TURBIDIMETER

 

 

 

 

 

Clean instrument housing                                                                     

M

 

 

Clean cells                                                                                             

D

 

 

 

 

CONDUCTIVITY METER

 

 

 

 

 

Check batteries and probe cables                                                        

D

 

 

Replatinize Probe                                                                                  

5

 

 

 

 

DISSOLVED OXYGEN METERS

 

 

 

 

PROBE

 

 

 

 

Check membrane for deterioration; check filling solution

D(1)

 

METER

 

 

 

 

Battery level and electronics checked                                                

D(1)

 

 

 

 

THERMOMETERS

 

 

 

 

 

Check for cracks and gaps in the mercury                                           

D(1)

 

 

 

 

TEMPERATURE PROBE

 

 

 

 

 

Check connections, cables                                                                   

D

 

 

Check against calibrated thermometer                                                 

D

 

 

 

 

AUTOSAMPLERS

 

 

 

 

 

Check needles and tubing                                                                    

D(1)

 

 

Clean                                                                                                     

Q


Table 8.1, cont.

PREVENTIVE MAINTENANCE ACTIVITIES

 

INSTRUMENT/ACTIVITY

 

 

FREQUENCY

 

DATA SYSTEMS

 

 

 

 

 

 

Clean computers, check battery backup and check                             

Q

 

 

 

 ventilation fans

 

 

 

 

 

 

 

AUTOMATIC SAMPLE COLLECTION SYSTEMS (ex. ISCO, Sigma, etc.)

 

 

 

 

 

Check sampler operation (forward, reverse, automatic                        

D(6)

 

 

 

 through three cycles of the purge-pump-purge cycle)

 

 

 

 

Check purge-pump-purge cycle when sampler is installed                   

D(7)

 

 

 

Check the flow pacer that activates the sampler to assure proper operation

 

 

 

 

Check desiccant                                                                                    

D(1,6)

 

 

 

Check batteries                                                                                     

D(1,6)

 

 

 

Check pumping rate against manufacturer's specifications

D(1,6)

 

 

KEY:

1

Replace as necessary

 

D

daily*

2

High background

 

W

weekly

3

Loss of sensitivity or failing resolution

 

M

monthly

4

Erratic response

 

Q

quarterly

5

QC failure

 

SA

semi-annually

6

Prior to sampling event

 

A

annually

7

In situ (under field conditions)

 

 

 

 

*Daily is defined as prior to use or a 12-hour period if equipment is run continuously